PARKER BATH AL21000-GB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2018-05-16 for PARKER BATH AL21000-GB manufactured by Arjohuntleigh Polska Sp. Z O.o..

Event Text Entries

[129312935] (b)(4). Please note that previous medwatch reports for this product may have been submitted for the manufacturing site arjo hospital equipment ab (under registration #(b)(4)). As of 2014 that number was de-activated due to the site no longer being a manufacturer and shipping product to the usa. From 2014 and going forward complaints related to these products are to be handled by arjohuntleigh ab's complaint handling establishment and any medwatch reports will be submitted under registration #(b)(4). Additional information will be provided within the next report.
Patient Sequence No: 1, Text Type: N, H10


[129312936] It was reported that caregiver was bathing the patient, when the door opened and the water flooded out. The caregiver caught the patient to stop her falling out of the bath. The patient did not sustain injury. The caregiver jarred her shoulder, when catching the patient. The caregiver taken pain killers. No malfunction of the bath was detected. According to the collected information the door was not locked fully and opened while the caregiver was on the other side of the bath.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2018-00225
MDR Report Key7517579
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2018-05-16
Date of Report2018-12-18
Date of Event2018-11-08
Date Facility Aware2018-11-08
Report Date2018-12-18
Date Reported to FDA2018-12-18
Date Reported to Mfgr2018-12-18
Date Mfgr Received2018-11-08
Device Manufacturer Date2016-12-19
Date Added to Maude2018-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KINGA STOLINSKA
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal62-052
Manufacturer G1ARJOHUNTLEIGH POLSKA SP. Z O.O.
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal Code62-052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePARKER BATH
Generic NameBATH, HYDRO-MASSAGE
Product CodeILJ
Date Received2018-05-16
Model NumberAL21000-GB
Device AvailabilityY
Device Age2 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARJOHUNTLEIGH POLSKA SP. Z O.O.
Manufacturer AddressKS. WAWRZYNIAKA 2 KOMORNIKI, 62-052 PL 62-052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-05-16