SHIELDED TOTALVIEW ENDOILLUMINATION PROBE INCL ILLUMINATED SCLERAL DEPRESSOR 25G 3269.SBS05

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-05-16 for SHIELDED TOTALVIEW ENDOILLUMINATION PROBE INCL ILLUMINATED SCLERAL DEPRESSOR 25G 3269.SBS05 manufactured by Dutch Opthalmic Research Center Bv.

Event Text Entries

[108900404] The investigation showed that when the fiber is inserted several times this kind of fracture is not possible. The investigation showed that the outside layer of the fiber is torn apart. This is not a production related error. This is a user error. Assuming worst case during surgery, the shaft tip got broken several times when the surgeon inserting the instrument into the patient's eye. This may cause a severe harm in the eye of the patient. Hence considered reportable. There is no information reasonably suggesting that the patient was harmed. This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)). Details of this activity were discussed with cdrh office of compliance (b)(4) during a tele-conference on (b)(4) 2017. All available information has been disclosed. This complaint has now been closed.
Patient Sequence No: 1, Text Type: N, H10


[108900405] During operation, the doctor inserted this instrument into a patient's eye several times. And then, the shaft tip part got broken. No further information is available. All information has been disclosed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222074-2018-00135
MDR Report Key7520326
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-05-16
Date of Report2015-10-22
Date of Event2015-10-22
Date Mfgr Received2015-10-22
Date Added to Maude2018-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS LAURA SMITH
Manufacturer Street10 CONTINENTAL DR BLDG 1
Manufacturer CityEXETER NH 03833
Manufacturer CountryUS
Manufacturer Postal03833
Manufacturer Phone6037786929
Manufacturer G1DUTCH OPTHALMIC RESEARCH CENTER BV
Manufacturer StreetSCHEIJDELVEWEG 2
Manufacturer CityZUIDLAND, 3214VN
Manufacturer CountryNL
Manufacturer Postal Code3214VN
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSHIELDED TOTALVIEW ENDOILLUMINATION PROBE INCL ILLUMINATED SCLERAL DEPRESSOR 25G
Generic NameSHIELDED TOTALVIEW ENDOILLUMINATION PROBE INCL ILLUMINATED SCLERAL DEPRESSOR 25G
Product CodeMPA
Date Received2018-05-16
Catalog Number3269.SBS05
Lot Number2000369609
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDUTCH OPTHALMIC RESEARCH CENTER BV
Manufacturer AddressSCHEIJDELVEWEG 2 ZUIDLAND, 3214VN NL 3214VN


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.