MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2018-05-17 for ORS-300N WARMER DRAPE manufactured by Microtek Dominicana, S.a..
[108615015]
No sample was received for evaluation. The device history record was reviewed. This lot had (b)(4) pieces that were manufactured from 09/18/17 to 09/19/17. No defects were reported during quality inspections based on the device history record, this does not appear to be the result of a personnel, process or material issue. Melts, though uncommon, can occur when the warmer is not draped properly, not turned off by using the power button, and/or there is insufficient fluid in the warmer basin, contrary to the operations manual, product labeling, product insert and in-service presentations. No further actions are being taken at this time.
Patient Sequence No: 1, Text Type: N, H10
[108615016]
Customer reports "we discovered a hole in the warmer drape ((b)(4)) provided in a craniotomy pack today" no patient injury or treatment was reported for the incident. This drape is sold non-sterile to a kit packer to be included in a procedure kit and sterilized before use.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8043817-2018-00018 |
MDR Report Key | 7523908 |
Report Source | DISTRIBUTOR,HEALTH PROFESSION |
Date Received | 2018-05-17 |
Date of Report | 2018-04-24 |
Date of Event | 2018-04-17 |
Date Mfgr Received | 2018-04-24 |
Date Added to Maude | 2018-05-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS JENNIFER MRKVICKA |
Manufacturer Street | 1 ECOLAB PLACE |
Manufacturer City | ST. PAUL MN 55102 |
Manufacturer Country | US |
Manufacturer Postal | 55102 |
Manufacturer Phone | 6512502237 |
Manufacturer G1 | MICROTEK DOMINICANA |
Manufacturer Street | ZONA FRANCA NO. 2 |
Manufacturer City | LA ROMANA, |
Manufacturer Country | DR |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ORS-300N WARMER DRAPE |
Generic Name | FLUID WARMING AND SLUSH DRAPES |
Product Code | LHC |
Date Received | 2018-05-17 |
Model Number | ORS-300N |
Lot Number | D172599 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MICROTEK DOMINICANA, S.A. |
Manufacturer Address | ZONA FRANCA NO. 2 LA ROMANA, DR |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-05-17 |