ORS-300N WARMER DRAPE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2018-05-17 for ORS-300N WARMER DRAPE manufactured by Microtek Dominicana, S.a..

Event Text Entries

[108615015] No sample was received for evaluation. The device history record was reviewed. This lot had (b)(4) pieces that were manufactured from 09/18/17 to 09/19/17. No defects were reported during quality inspections based on the device history record, this does not appear to be the result of a personnel, process or material issue. Melts, though uncommon, can occur when the warmer is not draped properly, not turned off by using the power button, and/or there is insufficient fluid in the warmer basin, contrary to the operations manual, product labeling, product insert and in-service presentations. No further actions are being taken at this time.
Patient Sequence No: 1, Text Type: N, H10


[108615016] Customer reports "we discovered a hole in the warmer drape ((b)(4)) provided in a craniotomy pack today" no patient injury or treatment was reported for the incident. This drape is sold non-sterile to a kit packer to be included in a procedure kit and sterilized before use.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2018-00018
MDR Report Key7523908
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2018-05-17
Date of Report2018-04-24
Date of Event2018-04-17
Date Mfgr Received2018-04-24
Date Added to Maude2018-05-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORS-300N WARMER DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-05-17
Model NumberORS-300N
Lot NumberD172599
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.