ETHANOL GEN.2 05967104190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-05-18 for ETHANOL GEN.2 05967104190 manufactured by Roche Diagnostics.

Event Text Entries

[108640538] This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[108640539] The customer stated that they received erroneous results for one patient sample tested for etoh2 ethanol gen. 2 (etoh) on an unknown roche analyzer. The sample initially resulted with an etoh value of 5. 06 g/l. The sample was automatically repeated, resulting as 5. 25 g/l. An erroneous result was reported outside of the laboratory to medical personnel. The doctor did not believe the values since they did not meet the clinical picture of the patient. No adverse events were alleged to have occurred with the patient. The model and serial number of the roche analyzer used were asked for, but not provided. The investigation was unable to find a definitive root cause. Product labeling for the etoh reagent instructs the customer to not use the automatic repeat function for repeating samples.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01543
MDR Report Key7525586
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-18
Date of Report2018-05-18
Date of Event2018-03-07
Date Mfgr Received2018-04-30
Date Added to Maude2018-05-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameETHANOL GEN.2
Generic NameALCOHOL DEHYDROGENASE, SPECIFIC REAGENT FOR ETHANOL ENZYME METHOD
Product CodeDML
Date Received2018-05-18
Model NumberNA
Catalog Number05967104190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-18