[109236691] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[109236692] It was reported to gore that the physician was implanting a gore? Viatorr? Tips endoprosthesis for a treatment of the portal vein hypertension due to the cirrhosis. The device was advanced through a 10fr cook rups100 sheath via a super stiff guide wire. After the device was well positioned, the physician pulled back the sheath to deploy the chainlink portion of device, then started initiating deployment of lined region. However when the lined region was deployed half, a resistance was reportedly felt and the deployment was not able to continue. The physician exchanged the 10fr sheath with an 11fr sheath then retracted the partially deployed viatorr? Device out of the patient through the 11fr sheath, there was no reported issue or injury to the patient during the device retraction. The physician used another viatorr? Device of same size to continue the procedure. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007284313-2018-00158
MDR Report Key7528976
Date Received2018-05-21
Date of Report2018-05-24
Date of Event2018-04-18
Device Manufacturer Date2017-09-25
Date Added to Maude2018-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactHELEN HUANG
Manufacturer Street1500 N. 4TH STREET
Manufacturer CityFLAGSTAFF AZ
Manufacturer Phone9285263030
Manufacturer G1MEDICAL PHOENIX 1 B/P
Manufacturer Street32360 N. NORTH VALLEY PARKWAY
Manufacturer CityPHOENIX AZ 85085
Manufacturer CountryUS
Manufacturer Postal Code85085
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameGORE? VIATORR? TIPS ENDOPROSTHESIS
Generic NameSHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Product CodeMIR
Date Received2018-05-21
Returned To Mfg2018-05-10
Catalog NumberPTB086275W
Lot Number17268183
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerW.L. GORE & ASSOCIATES
Manufacturer AddressFLAGSTAFF AZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-05-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.