MECHANICAL CHAIR/TRANSPORT CHAIR 547FL06-TS 54 SERIES RECLINER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2018-05-22 for MECHANICAL CHAIR/TRANSPORT CHAIR 547FL06-TS 54 SERIES RECLINER manufactured by Champion Manufacturing.

Event Text Entries

[108905360] Review of operator manual that accompanied the chair page 7 states proper in-service procedures. Pages 16 and 17 contain a caution that states: caution: this chair has moving parts that may create pinch points. Work carefully. Have your hands clear of the mechanism when you move the chair from one position to another. In addition to the initial shipment of chairs, there were four additional shipments of similar chairs received at this facility since this specific chair was purchased. All had operator manuals noting the proper operation of the recliners, including directions for staff to safely assist a patient in positioning the chair if they need assistance. The manual also contains cautions about keeping your hands clear of the mechanism. The in-service packet also comes with a video showing safe operation of the chair by a patient and by staff assisting the patient. Review of service parts and sales activity at this particular clinic shows no significant repair to the mechanical parts of their chairs. Routine follow up call noted chairs operating properly and there were no questions or concerns from staff. On 11 april 2018, champion verified with the facility that they no longer had serial (b)(4). The facility representative said he thought they discarded the chair when they purchased all new chairs from champion.
Patient Sequence No: 1, Text Type: N, H10


[108905361] A technician was assisting a patient onto a recliner. The actuator handle was pulled to release the footrest and the technician's hand was in the path of the opening footrest. The technician pinched his left ring finger in the mechanism of the chair, resulting in a skin trauma of laceration 2 cm square in size.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1834066-2018-00001
MDR Report Key7533699
Report SourceOTHER
Date Received2018-05-22
Date of Report2018-05-10
Date of Event2017-02-18
Date Mfgr Received2018-04-10
Device Manufacturer Date2004-12-01
Date Added to Maude2018-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationATTORNEY
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINE MILLER
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal46516
Manufacturer G1CHAMPION MANUFACTURING
Manufacturer Street2601 INDUSTRIAL PARKWAY
Manufacturer CityELKHART IN 46516
Manufacturer CountryUS
Manufacturer Postal Code46516
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMECHANICAL CHAIR/TRANSPORT CHAIR
Generic NameMECICAL RECLINER
Product CodeKMN
Date Received2018-05-22
Model Number547FL06-TS
Catalog Number54 SERIES RECLINER
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHAMPION MANUFACTURING
Manufacturer Address2601 INDUSTRIAL PARKWAY ELKHART IN 46516 US 46516


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-05-22