STATIC AND DYNAMIC COMPLIANCE PHYSICIANS CHOICE ENCORE 777544-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-05-23 for STATIC AND DYNAMIC COMPLIANCE PHYSICIANS CHOICE ENCORE 777544-102 manufactured by Vyaire Medical, Inc.

Event Text Entries

[109105936] (b)(4). As the correction for this event, the vyaire medial field service engineer installed a replacement module during an onsite visit. Results of investigation: the suspected device was received and a failure investigation was performed. The reported issue was confirmed and a burnt analyzer board was found. Upon further investigation, the failure analysis technician found there was shorted to ground by the pneumatic control board assembly mounting hardware. Several attempts have been made to contact the customer for additional information. At this time, the customer has not responded to any contact attempts and vyaire medical is unable to confirm the customer's reported device issue.
Patient Sequence No: 1, Text Type: N, H10


[109105937] The customer initially reported the static and dynamic compliance vmax device was showing lower than expected residual volume values.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2018-07897
MDR Report Key7539370
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-05-23
Date of Report2018-05-23
Date of Event2018-01-04
Date Mfgr Received2018-05-07
Device Manufacturer Date2011-08-23
Date Added to Maude2018-05-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICATED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2018-05-23
Returned To Mfg2018-01-17
Model NumberPHYSICIANS CHOICE ENCORE
Catalog Number777544-102
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.