30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-24 for 30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250 manufactured by Charter Medical, Ltd..

Event Text Entries

[109518072] The complaint sample, or pictures of the complaint device, was not returned to the manufacturer for investigation/review. The reported leak did not occur during filling of the bag, which suggests the leak likely occurred during handling after the bag was frozen. There was no additional information to suggest an exact root cause for this report. There are several potential issues which could contribute to a failure mode of this nature: residual moisture on the outside of the bag when it was placed inside the cassette. Residual moisture could cause the bag to freeze/adhere to the cassette during the freezing process. A condition of this type could cause film damage upon removal of a frozen bag from the cassette. Handling of the bag in the frozen state. The film of a frozen bag is fragile - inadvertent manipulation or impact on the bag could cause the film to fracture. The precautions indicated above are addressed in the product ifu's.
Patient Sequence No: 1, Text Type: N, H10


[109518073] The end user experienced a leak from a possible film tear in a cf-250 device while the bag was being thawed. The cell product was infused into the patient. A sterility sample was also taken from the cell product in the bag. The sterility test was negative for growth.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00010
MDR Report Key7542103
Date Received2018-05-24
Date of Report2018-05-24
Date of Event2018-04-24
Date Mfgr Received2018-04-24
Device Manufacturer Date2017-12-02
Date Added to Maude2018-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-05-24
Model NumberCF-250
Catalog NumberCF-250
Lot Number151496
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.