LIFECODES HLA-DQA1/B1 SSO TYPING KIT 628930

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-05-24 for LIFECODES HLA-DQA1/B1 SSO TYPING KIT 628930 manufactured by Immucor Gti Diagnostics, Inc..

Event Text Entries

[109627091] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[109627092] Immucor received a customer complaint on 26 february 2018 ((b)(4)) from a customer that contacted immucor regarding eight samples across three runs that generated a mistype with lifecodes hla-dqa1/b1 sso typing kit (lot 3006117). The customer reports that for these samples the dqa probe 207 was false negative resulting in dqa sample mistype for dqa1*02. In all cases, the samples were typed as dqa1*01 homozygous by lifecodes hla-dqa1/dqb1 sso typing kit lot 3006117. The samples were repeated for dqa locus by the customer on ssp/linkseq and the repeat typing results were dqa1*01, dqa1*02. The customer informed immucor that no incorrect results were reported. The customer chose to repeat the typings by another method due to probe 207 being close to the threshold and due to known gene associations with drb1.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183608-2018-00010
MDR Report Key7542134
Report SourceUSER FACILITY
Date Received2018-05-24
Date of Report2018-02-26
Date of Event2018-02-15
Date Mfgr Received2018-02-26
Device Manufacturer Date2017-12-20
Date Added to Maude2018-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JULIE EWEND
Manufacturer Street20925 CROSSROADS CIRCLE
Manufacturer CityWAUKESHA WI 53186
Manufacturer CountryUS
Manufacturer Postal53186
Manufacturer Phone2627541009
Manufacturer G1IMMUCOR GTI DIAGNOSTICS, INC.
Manufacturer Street20925 CROSSROADS CIRCLE
Manufacturer CityWAUKESHA WI 53186
Manufacturer CountryUS
Manufacturer Postal Code53186
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIFECODES HLA-DQA1/B1 SSO TYPING KIT
Generic NameLIFECODES HLA-DQA1/B1 SSO TYPING KIT
Product CodeMZI
Date Received2018-05-24
Catalog Number628930
Lot Number3006117
Device Expiration Date2019-06-15
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR GTI DIAGNOSTICS, INC.
Manufacturer Address20925 CROSSROADS CIRCLE WAUKESHA WI 53186 US 53186


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-24

© 2026 FDA.report