[109753373] (b)(4). (b)(6). Device requested for investigation but was unavailable.
Patient Sequence No: 1, Text Type: N, H10


[109753374] The customer received questionable cholesterol results from accutrend plus meter serial (b)(4). The date of event is an approximation. The customer provided cholesterol results of "lo" (< 150 mg/dl), 250 mg/dl, and 165 mg/dl. Different fingers were used for each measurement. It was not known when the "lo" result was obtained, but it was provided that the 250 mg/dl and the 165 mg/dl results were obtained 15 minutes apart. It is unknown if the erroneous results were released to the patient or the patient's doctor. There was no allegation of an adverse event. It was provided that the customer did not apply enough blood to cover the entire test area. The customer's current condition was provided as well. The customer's meter and strips were requested for investigation. The customer's strips were no longer available. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01594
MDR Report Key7544153
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-25
Date of Report2018-05-31
Date of Event2018-05-07
Date Mfgr Received2018-05-07
Date Added to Maude2018-05-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT FAMILY MEMBER OR FRIEND
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCUTREND CHOLESTEROL
Generic NameCHOLESTEROL (TOTAL) TEST SYSTEM
Product CodeCHH
Date Received2018-05-25
Model NumberNA
Catalog Number11418262171
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-25