CAPTURE-R READY INDICATOR RED CELLS 0006428

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-05-25 for CAPTURE-R READY INDICATOR RED CELLS 0006428 manufactured by Immucor, Inc..

Event Text Entries

[109747218] The full number for the product in question is bk020053 (1/27/03). Immucor technical support used a remote electronic connection method to assess the instrument test well images in question on 01may2018, which visually appeared positive. A review of the check image well showed a bubble in the 3rd screening well, and the event log showed that there was a warning for the reagent in question of a remaining liquid volume lower than expected. The immucor laboratory tested retention product on 17may2018 and 21may2018 against antibody screen strip r969, which performed as expected. Complaints of positive reactions being interpreted as negative by the galileo echo camera algorithm are being corrected under design control project (b)(4). The echo instrument is performing as expected with the exception of the camera module ability to appropriately identify some true positive results. The immucor technical communication (b)(4) is being used by the lab to visually confirm all negative assay events reported by the echo instrument. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[109747219] On (b)(6) 2018, a customer site reported an unexpected negative antibody screen when using capture-r ready indicator red cells by galileo echo.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1034569-2018-00127
MDR Report Key7545105
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-05-25
Date of Report2018-05-25
Date of Event2018-05-01
Date Mfgr Received2018-05-01
Date Added to Maude2018-05-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR HOWARD YOREK
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal Code300915625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R READY INDICATOR RED CELLS
Generic NameANTIGLOBULIN COATED RED CELLS
Product CodeKSF
Date Received2018-05-25
Catalog Number0006428
Lot Number221081
Device Expiration Date2018-05-20
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 300915625 US 300915625


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.