HUT EXT DR FINAL ASSY-STANDARD 404008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-29 for HUT EXT DR FINAL ASSY-STANDARD 404008 manufactured by Liebel-flarsheim.

Event Text Entries

[109757314] Table is shutting off during cases. Customer states generator screen in the console would go offline as if it lost power, could go over to on/off switch and turn on and it would turn right back on. The 2 monitors on the table were also going black, and could not fluoro. Only way to get back on was to reboot the generator. But today, it was working fine. Main issue - the console/generator is spontaneously shut down, no errors in the log. Nothing per ge fse. Fse reseated several connectors, power relay, etc, and it still shut off. Procedure was not completed and patient was moved to ct and finished. No injuries to patient or staff reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1518293-2018-00016
MDR Report Key7549347
Date Received2018-05-29
Date of Report2018-05-29
Date of Event2018-05-02
Date Mfgr Received2018-05-02
Device Manufacturer Date2008-05-31
Date Added to Maude2018-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFRED RECKELHOFF
Manufacturer Street2111 E. GALBRAITH RD
Manufacturer CityOH 45237
Manufacturer CountryUS
Manufacturer Postal45237
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHUT EXT DR FINAL ASSY-STANDARD
Generic NameHUT EXT DR FINAL ASSY-STANDARD
Product CodeIXR
Date Received2018-05-29
Model Number404008
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerLIEBEL-FLARSHEIM
Manufacturer Address2111 E. GALBRAITH RD CINCINNATI OH 45237 US 45237


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-29

© 2026 FDA.report