MALECOT RUSSEL GASTROSTOMY CATHETER ULT10.2-NT-23-P-NS-

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,08 report with the FDA on 2005-11-11 for MALECOT RUSSEL GASTROSTOMY CATHETER ULT10.2-NT-23-P-NS- manufactured by Cook, Inc..

Event Text Entries

[16830939] Catheter was placed in an outpt in 2005. It was spontaneously exited on the same day. The physician felt this situation occurred due to the softer material of the tubing. A replacement was used.
Patient Sequence No: 1, Text Type: D, B5


[16986100] Evaluation: the device was returned in a used condition. An examination found the inner catheter slightly elongated near the black indicator line. During testing, the inner catheter separated from its bonded position at the distal tip easily when pulled. An examination of this area found no evidence that the tubing was properly bonded at time of manufacture.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1820334-2005-00296
MDR Report Key755673
Report Source01,05,08
Date Received2005-11-11
Date of Report2005-10-13
Date of Event2005-09-28
Date Facility Aware2005-09-28
Report Date2005-10-13
Date Mfgr Received2005-10-13
Date Added to Maude2006-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactRITA HARDEN
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMALECOT RUSSEL GASTROSTOMY CATHETER
Generic NameCATHETER
Product CodeEZK
Date Received2005-11-11
Returned To Mfg2005-11-01
Model NumberNA
Catalog NumberULT10.2-NT-23-P-NS-
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeUNKNOWN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key743511
ManufacturerCOOK, INC.
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US
Baseline Brand NameMALECOT RUSSEL GASTROSTOMY CATHETER
Baseline Generic NameCATHETER
Baseline Model NoNA
Baseline Catalog NoULT10.2-NT-23-P-NS-
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2005-11-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.