MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-01 for ISPAN GAS TANK REGULATOR 8065797303 manufactured by Air Liquide.
[110043441]
No sample has been received by manufacturing for evaluation. Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[110043442]
A biomedical engineer reported that an ophthalmic gas dispensing regulator valve would not allow gas to outflow when the adjustment knob was turned. Procedure details, patient involvement and patient impact information is unknown. Additional information has been requested. Additional information received confirmed that this event was noted prior to the start of surgery. There was no patient involvement and no patient impact associated with this reported event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1610287-2018-00019 |
MDR Report Key | 7562553 |
Date Received | 2018-06-01 |
Date of Report | 2018-10-03 |
Date of Event | 2018-05-22 |
Date Mfgr Received | 2018-09-20 |
Device Manufacturer Date | 2017-12-11 |
Date Added to Maude | 2018-06-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. BRYAN BLAKE |
Manufacturer Street | 6201 SOUTH FREEWAY MAIL STOP AB2-6 |
Manufacturer City | FORT WORTH TX 76134 |
Manufacturer Country | US |
Manufacturer Postal | 76134 |
Manufacturer Phone | 8176152230 |
Manufacturer G1 | AIR LIQUIDE |
Manufacturer Street | 13140 TI BLVD. |
Manufacturer City | DALLAS TX 75243 |
Manufacturer Country | US |
Manufacturer Postal Code | 75243 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 0 |
Brand Name | ISPAN GAS TANK REGULATOR |
Generic Name | INTRAOCULAR GAS |
Product Code | LPO |
Date Received | 2018-06-01 |
Returned To Mfg | 2018-09-19 |
Model Number | NA |
Catalog Number | 8065797303 |
Lot Number | 734504 |
ID Number | 380657973033 |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | AIR LIQUIDE |
Manufacturer Address | 13140 TI BLVD. DALLAS TX 75243 US 75243 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2018-06-01 |