OLYMPIC COOL-CAP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-06-01 for OLYMPIC COOL-CAP manufactured by Natus Medical Inc..

Event Text Entries

[110365242] The asc used certified calibrated instruments, and it was determined that the temperature sensors work correctly. At the same time, a flow thermometer was used to verify the temperature of the water flow. The asc found that the machine is not able to get to the temperature of 15? C. The asc performed testing per natus service manual instructions. The asc believes the cooling unit has failed and has contacted technical service for repair options. An email requesting more information was sent to the customer on 5/25/2018, and the customer has not responded.
Patient Sequence No: 1, Text Type: N, H10


[110365243] An end user reported to the authorized service center (asc) that the cool cap temperature readings were suspected to be incorrect. There has been no report of a death or serious injury that required treatment, a delay in treatment, and an environmental or safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00131
MDR Report Key7562963
Report SourceUSER FACILITY
Date Received2018-06-01
Date of Report2018-05-03
Date Mfgr Received2018-05-03
Date Added to Maude2018-06-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-06-01
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC.
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-01

© 2026 FDA.report