CLEARVIEW STREP A EXACT II DIPSTICK 30T IST-501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2018-06-01 for CLEARVIEW STREP A EXACT II DIPSTICK 30T IST-501 manufactured by Alere San Diego, Inc.

Event Text Entries

[109898568] Investigation conclusion: false negative strep a results were not replicated in-house with retain products. Retain products were tested with strep a cutoff positive sample (2. 5e07org/ml). All results were positive at read time and met qc specifications. No false negative results were obtained during in-house testing. Manufacturing batch record review did not uncover any abnormalities. A negative result may be obtained if the concentration of the strep a antigen present in the throat swab is not adequate or is below the detectable level of the test. Based on the information available, there is no indication of a product deficiency and no corrective action is required.
Patient Sequence No: 1, Text Type: N, H10


[109898569] On (b)(6) 2018, the patient presented at the urgent care facility with a sore throat. Upon the initial examination, the customer noted that the patient's throat only appeared a little red and it was not immediately evident the patient had strep a. A test was performed with a clearview strep a exact ii dipstick test using a dual swab technique and a negative result was obtained at the 5 minute read time. The second throat swab was sent to (b)(6) laboratory for culture confirmation. The patient was diagnosed with pharyngitis and sent home. Two hours after test was performed on the clearview strep a exact ii dipstick test, the operator noticed a faint positive line had developed in the test region and forwarded this information to the customer. On (b)(6) 2018, the customer contacted the patient to inform the patient of the positive result and prescribe amoxicillin. At this time, the throat culture results had not been received. On (b)(6) 2018, the customer obtained a positive result from the confirmatory throat culture. The report indicated that there were a few colonies of strep a on the culture. The culture was performed on strep a specific bacitracin disk. The customer proceeded to attempt to contact patient to inform of the positive confirmatory test and ensure the patient had begun the prescribed course of antibiotics. On (b)(6) 2018, the patient returned to the urgent care facility for an unrelated issue and was provided the results of the confirmatory throat culture. The patient subsequently began the prescribed amoxicillin. Troubleshooting was conducted with the customer. The customer was advised results should not be interpreted after the read time and a faint positive line after the read time may indicate the sample concentration is below the level of detection.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2018-00068
MDR Report Key7563182
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2018-06-01
Date of Report2018-06-01
Date of Event2018-04-27
Date Mfgr Received2018-05-02
Date Added to Maude2018-06-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLEARVIEW STREP A EXACT II DIPSTICK 30T
Generic NameSTREP A RAPID TEST STRIP
Product CodeGTY
Date Received2018-06-01
Model NumberIST-501
Lot NumberSTA7110016
Device Expiration Date2019-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-01