FREEDOM CYCLER PD+ NEWTON IQ NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2005-11-10 for FREEDOM CYCLER PD+ NEWTON IQ NA manufactured by Fresenius Medical Care North America.

Event Text Entries

[539127] A home peritoneal dialysis patient reported that the cycler was skipping to fill 1 without completing drain 0. The patient has a last fill of 2,000 ml. He did not have the details, i. E. , drain volumes. The cycler was replaced and the problem has not recurred. There was not enough information during initial contract to make an mdr determination. The cycler was received by the manufacturer on october 14, 2005. The data sheets which were retrived on october 27, 2005 show a drain 0 time of 2 minutes and a drain 0 volume of 30. Fill 1 was 2,000 ml. And drain 1 was 4,600 ml.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2005-00020
MDR Report Key756474
Report Source04
Date Received2005-11-10
Date of Report2005-09-30
Date of Event2005-09-30
Report Date2005-09-30
Date Mfgr Received2005-09-30
Device Manufacturer Date2002-09-01
Date Added to Maude2006-09-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELVY DIZON, BSN, RN, CNN
Manufacturer Street2637 SHADELANDS DRIVE
Manufacturer CityWALNUT CREEK CA 94598
Manufacturer CountryUS
Manufacturer Postal94598
Manufacturer Phone*
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFREEDOM CYCLER PD+
Generic NamePERITONEAL DIALYSIS CYCLER
Product CodeKPF
Date Received2005-11-10
Returned To Mfg2005-10-14
Model NumberNEWTON IQ
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Age3 YR
Device Eval'ed by MfgrN
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key744358
ManufacturerFRESENIUS MEDICAL CARE NORTH AMERICA
Manufacturer Address* WALNUT CREEK CA * US
Baseline Brand NameFREEDOM CYCLER PD+
Baseline Generic NamePERITONEAL DIALYSIS CYCLER
Baseline Model NoNEWTON IQ
Baseline Catalog No180-60300 (NEW)
Baseline IDNONE
Baseline Device FamilyNA
Baseline Shelf Life ContainedN
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK915634
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2005-11-10