??SHOCKPUL SE?LITHOTRIPSY?PROBE, STERILE?SINGLE?USE,?3.76MM?(3/BOX) ??SPL-PD376

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-04 for ??SHOCKPUL SE?LITHOTRIPSY?PROBE, STERILE?SINGLE?USE,?3.76MM?(3/BOX) ??SPL-PD376 manufactured by Cybersonics Inc..

Event Text Entries

[110181268] The pouch of the subject device is requested from the customer. Following additional information is shared by the customer: "one person carried the box into the operating room and pulled out one probe from the box and discovered it was open at the top by the peel area. The other 2 were inspected from the box and were intact. No pictures are available right now. " ifu instructs to inspect the product/package for any damage. "do not use if package is damaged" symbol is on label of the product and also in the ifu. If significant additional information is received, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[110181269] "spl-pd376 sterile packaging was open upon opening outer box. No patient involvement. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004216443-2018-00001
MDR Report Key7564805
Date Received2018-06-04
Date of Report2018-06-01
Date of Event2018-05-16
Date Mfgr Received2018-05-22
Device Manufacturer Date2018-02-05
Date Added to Maude2018-06-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SAMRADNI PATIL
Manufacturer Street5340 FRYLING ROAD SUITE 101
Manufacturer CityERIE PA 165104703
Manufacturer CountryUS
Manufacturer Postal165104703
Manufacturer Phone8149201510
Manufacturer G1CYBERSONICS INC.
Manufacturer Street5340 FRYLING ROAD SUITE 101
Manufacturer CityERIE PA 165104703
Manufacturer CountryUS
Manufacturer Postal Code165104703
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name??SHOCKPUL SE?LITHOTRIPSY?PROBE, STERILE?SINGLE?USE,?3.76MM?(3/BOX)
Generic NameELECTROHYDRAULIC LITHOTRIPTOR
Product CodeFEO
Date Received2018-06-04
Model Number??SPL-PD376
Lot NumberD1800854
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCYBERSONICS INC.
Manufacturer Address5340 FRYLING ROAD SUITE 101 ERIE PA 165104703 US 165104703


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-04