[110043146] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[110043147] The customer stated that they received questionable results for an unspecified number of patient samples tested for ck creatine kinase (ck) and gluc3 glucose hk (glu) on a cobas 6000 c (501) module - c501. The customer stated that the issue occurred when the reagent packs were nearly empty. The customer provided data for one patient sample with an erroneous glu result. The erroneous result was not reported outside of the laboratory. The sample initially resulted with a glu value of < 20 mg/dl. The reagent pack was replaced and after this, the sample was repeated, resulting with a value that was expected. No adverse events were alleged to have occurred with the patient. The c501 analyzer serial number was asked for, but not provided. The issue was resolved after replacement of the reagent pack. The investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01737
MDR Report Key7568545
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-06-05
Date of Report2018-06-05
Date of Event2018-05-17
Date Mfgr Received2018-05-17
Date Added to Maude2018-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameGLUC3 GLUCOSE HK GEN.3
Generic NameHEXOKINASE, GLUCOSE
Product CodeCFR
Date Received2018-06-05
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.