MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-05 for GLUC3 GLUCOSE HK GEN.3 manufactured by Roche Diagnostics.
[110043146]
(b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[110043147]
The customer stated that they received questionable results for an unspecified number of patient samples tested for ck creatine kinase (ck) and gluc3 glucose hk (glu) on a cobas 6000 c (501) module - c501. The customer stated that the issue occurred when the reagent packs were nearly empty. The customer provided data for one patient sample with an erroneous glu result. The erroneous result was not reported outside of the laboratory. The sample initially resulted with a glu value of < 20 mg/dl. The reagent pack was replaced and after this, the sample was repeated, resulting with a value that was expected. No adverse events were alleged to have occurred with the patient. The c501 analyzer serial number was asked for, but not provided. The issue was resolved after replacement of the reagent pack. The investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2018-01737 |
MDR Report Key | 7568545 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-06-05 |
Date of Report | 2018-06-05 |
Date of Event | 2018-05-17 |
Date Mfgr Received | 2018-05-17 |
Date Added to Maude | 2018-06-05 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GLUC3 GLUCOSE HK GEN.3 |
Generic Name | HEXOKINASE, GLUCOSE |
Product Code | CFR |
Date Received | 2018-06-05 |
Model Number | NA |
Catalog Number | ASKU |
Lot Number | ASKU |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-06-05 |