RUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS 04618793190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-05 for RUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS 04618793190 manufactured by Roche Diagnostics.

Event Text Entries

[110125335] Unique identifier (udi)# (b)(4). The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[110125336] The customer complained of questionable elecsys rubella igg immunoassay (rubella igg) results for 2 patient samples tested on a cobas 8000 e 602 module. For patient 1 the initial rubella igg result was 8 ui/ml, non reactive. The rubella igg result from another lot of reagent was 16 ui/ml, reactive. For patient 2 the initial rubella igg result was 8 ui/ml, non reactive. The rubella igg result from another lot of reagent was 16 ui/ml, reactive. The lot information for the other lot of rubella igg reagent was requested but was not provided. It was unknown if the erroneous results were reported outside of the laboratory. There was no allegation of an adverse event. The cobas e602 serial number was (b)(4). The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01751
MDR Report Key7570407
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-06-05
Date of Report2018-08-23
Date of Event2018-04-20
Date Mfgr Received2018-05-18
Date Added to Maude2018-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameRUBELLA IGG - IGG ANTIBODIES TO RUBELLA VIRUS
Generic NameENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA
Product CodeLFX
Date Received2018-06-05
Model NumberNA
Catalog Number04618793190
Lot Number283664
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.