INTRATEMP THERMABASIN DRAPE ITB100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-06-05 for INTRATEMP THERMABASIN DRAPE ITB100 manufactured by Microtek Medical, Inc..

Event Text Entries

[110195427] One sample was received 5/18/18 in (b)(6). The sample was opened, red bagged and contaminated. Photos were taken to show the crack in the bottom of the basin. The sample is being discarded appropriately and will not be sent to (b)(6). A scar was issued to the supplier for this complaint. The investigation is pending a response from the supplier.
Patient Sequence No: 1, Text Type: N, H10


[110195428] End user reported basin has a crack in it. Medwatch form mw5077290 was received on 5/30/18 with the following information: ecolab intratemp warming unit used during surgery to repair an abdominal aortic aneurysm. Upon tear down of sterile field, it was noted that basin had a small crack in the underneath of basin. Crack was unable to be visualized from sterile portion. The physician was notified, infectious disease was consulted, baseline blood cultures were obtained and patient was placed on further antibiotic therapy for 2 weeks. The patient was stable without fever and was closely monitored. Blood cultures had no growth after 4 days. Pharmacy was dosing vancomycin.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3013401749-2018-00004
MDR Report Key7571298
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-06-05
Date of Report2018-05-11
Date of Event2018-05-10
Date Mfgr Received2018-05-11
Date Added to Maude2018-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK MEDICAL, INC.
Manufacturer Street13000 DEERFIELD PARKWAY SUITE 300
Manufacturer CityALPHARETTA GA 30004
Manufacturer CountryUS
Manufacturer Postal Code30004
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTRATEMP THERMABASIN DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-06-05
Returned To Mfg2018-05-18
Model NumberITB100
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK MEDICAL, INC.
Manufacturer Address13000 DEERFIELD PARKWAY SUITE 300 ALPHARETTA GA 30004 US 30004


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-05