DICKSON PARAFFIN BATH PB-107

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-05 for DICKSON PARAFFIN BATH PB-107 manufactured by Fabricacion Y Manufacturas De Mexico S.a. De C.v..

Event Text Entries

[110169931] Evaluation of the unit found that the chassis had been entered by an unknown third party. Wiring connectors and thermostat connections were found to be modified and not meeting device specifications. The failure was documented for trend analysis. Credit was issued to customer after return.
Patient Sequence No: 1, Text Type: N, H10


[110169932] Customer reported an issue where a unit got hot and burned a user. Multiple attempts to contact the distributer failed and therefore additional event details could not be obtained. The customer that contacted us did not clarify the severity of the burn.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2222003-2018-00002
MDR Report Key7571465
Date Received2018-06-05
Date of Report2018-02-27
Date of Event2014-03-22
Date Mfgr Received2014-03-28
Device Manufacturer Date2013-10-14
Date Added to Maude2018-06-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN KAHLE
Manufacturer Street15125 PROCTOR AVE
Manufacturer CityCITY OF INDUSTRY CA 91746
Manufacturer CountryUS
Manufacturer Postal91746
Manufacturer Phone6269686681
Manufacturer G1FABRICACION Y MANUFACTURAS DE MEXICO
Manufacturer StreetMAQUILADORAS 1387-1 COL. CIUDAD INDUSTRIAL
Manufacturer CityTIJUANA, B.C. 22444
Manufacturer CountryMX
Manufacturer Postal Code22444
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDICKSON PARAFFIN BATH
Generic NamePARAFFIN THERAPY
Product CodeIMC
Date Received2018-06-05
Model NumberPB-107
Catalog NumberPB-107
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerFABRICACION Y MANUFACTURAS DE MEXICO S.A. DE C.V.
Manufacturer AddressMAQUILADORAS 1387-1 COL.CIUDAD INDUSTRIAL TIJUANA, B.C. C.P. 22444 MX 22444


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.