DRAGER JAUNDICE METER, MODEL JM-103 MU20606

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-06 for DRAGER JAUNDICE METER, MODEL JM-103 MU20606 manufactured by Draeger Medical Systems, Inc..

Event Text Entries

[110231627]
Patient Sequence No: 1, Text Type: N, H10


[110231628] The meter (jm-103) has a value of 1- that indicates one more flash needs to be completed before the numbers can be averaged and the result is calculated. The problem is that this number has the mg/dl under it which makes it really seem like a result...... Extremely easy to misread and i'm surprised this doesn't happen more often. Granted, the second screen does have 2 dashes before the value and isn't in a decimal format, but still potential for error (obviously).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7573133
MDR Report Key7573133
Date Received2018-06-06
Date of Report2018-05-22
Date of Event2018-05-22
Report Date2018-05-22
Date Reported to FDA2018-05-22
Date Reported to Mfgr2018-05-22
Date Added to Maude2018-06-06
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRAGER JAUNDICE METER, MODEL JM-103
Generic NameBILIRUBIN METER
Product CodeMQM
Date Received2018-06-06
Model NumberMU20606
Catalog NumberJM-103
Lot NumberNA
OperatorNURSE
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDRAEGER MEDICAL SYSTEMS, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US 18969


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-06