CYSTOSCOPY IRRIGATION SET JC4040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-06-06 for CYSTOSCOPY IRRIGATION SET JC4040 manufactured by Baxter Healthcare - Cartago.

Event Text Entries

[110368131] (b)(4). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[110368132] It was reported that there was particulate matter in a cysto/bladder irrigation set. The reporter stated that after the packaging was opened and the medication spiked, a small piece of red plastic was noted in the chamber. Upon further inspection, the customer noticed that the red cap over the spike was split and a small piece was missing. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[114187903] The device was received for evaluation. A visual inspection was performed and a small piece of red plastic was observed inside the chamber or housing. A spectrophotometric comparison was performed on the particulate matter and it was determined to of the same material as a component. The reported condition was verified. The cause of the reported condition was an assembly issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[114187904]
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2018-03436
MDR Report Key7573241
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-06-06
Date of Report2018-07-13
Date Mfgr Received2018-07-12
Device Manufacturer Date2017-09-08
Date Added to Maude2018-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10 SEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCYSTOSCOPY IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2018-06-06
Returned To Mfg2018-07-10
Model NumberNA
Catalog NumberJC4040
Lot NumberR17I07023
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CARTAGO
Manufacturer AddressCARTAGO


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.