256 CHANNEL GES 400 CLINICAL EEG SYSTEM WITH IMAC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-06-06 for 256 CHANNEL GES 400 CLINICAL EEG SYSTEM WITH IMAC manufactured by Electrical Geodesics, Inc..

Event Text Entries

[110709608] The ges 400 system was being used in an emu room to record a patient's eeg. The ges 400 cart was positioned next to the patient 's bed with the camera parapet positioned above the patient. There were two emu staff near the cart, with the lead technician training a new technician. The technicians were adjusting the camera pole higher (pressed button to raise the pole) when they noticed that the camera pole became detached from the base as it was being raised. The lead technician asked the new technician to hold the pole to support the camera while she went to get the eeg department manager. When the manager came to the room, they assessed the situation and further movement caused the camera pole to mechanically detach from the cart as it was being held. The camera pole remained captive to the cart by way of the camera signal cable inside of the pole. The separation was controlled and no debris was released; there was no patient involvement. There was no allegation of patient or user injury. The eeg case was completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001090553-2018-00001
MDR Report Key7574942
Report SourceUSER FACILITY
Date Received2018-06-06
Date of Report2018-05-09
Date of Event2018-05-09
Date Mfgr Received2018-05-09
Device Manufacturer Date2017-10-20
Date Added to Maude2018-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NANCY ANNUNZIATO
Manufacturer Street500 EAST 4TH AVENUE SUITE 200
Manufacturer CityEUGENE OR 97401
Manufacturer CountryUS
Manufacturer Postal97401
Manufacturer Phone5034076470
Manufacturer G1ELECTRICAL GEODESICS INC.
Manufacturer Street500 EAST 4TH AVENUE SUITE 200
Manufacturer CityEUGENE OR 97401
Manufacturer CountryUS
Manufacturer Postal Code97401
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name256 CHANNEL GES 400 CLINICAL EEG SYSTEM WITH IMAC
Generic NameGES 400
Product CodeGWQ
Date Received2018-06-06
Returned To Mfg2018-05-22
Lot NumberZZ-01 3/29/16 8:29
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerELECTRICAL GEODESICS, INC.
Manufacturer Address500 EAST 4TH AVENUE SUITE 200 EUGENE OR 97401 US 97401


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-06

© 2026 FDA.report