ADSORBA 115263

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-06 for ADSORBA 115263 manufactured by Baxter Healthcare - Hechingen Baden Dialyzer.

Event Text Entries

[110715713] (b)(6). The actual samples were not available; however, photographs of the samples were provided for evaluation. During visual inspection of the provided photos, black particles in the tubing and in the header cap of the dialyzer were visible for both samples. The reported condition was verified. The cause of the condition was not determined. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[110715714] It was reported that black particles were observed in the priming solution of two (2) adsorba 150 c during priming. The dialyzers had been primed with 5% glucose solution and 0. 9% saline solution. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611369-2018-00058
MDR Report Key7575015
Date Received2018-06-06
Date of Report2018-07-07
Date of Event2018-05-03
Date Mfgr Received2018-07-03
Device Manufacturer Date2017-07-31
Date Added to Maude2018-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
Manufacturer StreetHOLGER - CRAFOORD - STRASSE 26
Manufacturer CityHECHINGEN D-72379
Manufacturer CountryGM
Manufacturer Postal CodeD-72379
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameADSORBA
Generic NameAPPARATUS, HEMOPERFUSION, SORBENT
Product CodeFLD
Date Received2018-06-06
Returned To Mfg2018-06-21
Model NumberNA
Catalog Number115263
Lot Number7-656
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
Manufacturer AddressHECHINGEN


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-06

© 2026 FDA.report