OPTICAL BIOPSY FORCEPS 27072A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2018-06-06 for OPTICAL BIOPSY FORCEPS 27072A manufactured by Karl Storz Se & Co. Kg.

Event Text Entries

[110583122] The device has not been returned, and we have been unable to obtain any further details about the event to determine a possible cause of the laceration. The picture of the instrument provided by the customer shows the jaws are intact, and the rivet/ hinge pin (approx. 1mm in size) is missing. If the pin (rivet) came off of the forceps during the procedure, this would not be likely to cause injury to the patient and the pin would most likely be passed when urinating. Per the lot code provided, the instrument had been in use for over 8 years. The most likely cause of the damage is wear of use.
Patient Sequence No: 1, Text Type: N, H10


[110583123] Allegedly, a rivet holding the biopsy cup jaw fell out during a urological procedure in which the patient sustained a minor injury (laceration), requiring a catheter until the laceration could heal.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610617-2018-00040
MDR Report Key7575514
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2018-06-06
Date of Report2018-05-07
Date of Event2018-05-03
Date Mfgr Received2018-05-07
Date Added to Maude2018-06-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO CA 902455017
Manufacturer CountryUS
Manufacturer Postal902455017
Manufacturer Phone4242188201
Manufacturer G1KARL STORZ SE & CO. KG
Manufacturer StreetDR.-KARL-STORZ-STRASSE 34 78532
Manufacturer CityTUTTLINGEN,
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTICAL BIOPSY FORCEPS
Generic NameOPTICAL BIOPSY FORCEPS
Product CodeOCZ
Date Received2018-06-06
Model Number27072A
Catalog Number27072A
Lot NumberNZ
OperatorPHYSICIAN
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ SE & CO. KG
Manufacturer AddressDR.-KARL-STORZ-STRASSE 34 78532 TUTTLINGEN, GM