BENEVISION CENTRAL MONITORING SYSTEM 115-034413-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-06-07 for BENEVISION CENTRAL MONITORING SYSTEM 115-034413-00 manufactured by Shenzhen Mindray Bio-medical Electronics Co., Ltd.

Event Text Entries

[110332367] The company service representative evaluated the system and confirmed the unit was generating visual and audible alarms. The company service representative increased the volume and locked the display with the authority of customer biomed present and a new password was reset. The customer is not alleging any device malfunction and reported that no additional information will be available. The operative manual provides a warning to remind the user of the risk of setting alarm volume to a low level. There was no device malfunction. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[110332368] The customer reported that the audio level in the display menu of the central monitoring station had been turned down and a patient being monitored on the system expired. Note: the customer does not allege that the system missed an alarm.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009156722-2018-00006
MDR Report Key7577371
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-06-07
Date of Report2018-05-14
Date of Event2018-05-12
Date Facility Aware2018-05-14
Report Date2018-06-04
Date Reported to Mfgr2018-06-04
Date Mfgr Received2018-05-14
Date Added to Maude2018-06-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SERENA CHEN
Manufacturer StreetKEJI 12TH ROAD SOUTH HI-TECH INDUSTRIAL PARK
Manufacturer CitySHENZHEN, NANSHAN 518057
Manufacturer CountryCH
Manufacturer Postal518057
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBENEVISION CENTRAL MONITORING SYSTEM
Generic NameBENEVISION
Product CodeMSX
Date Received2018-06-07
Model NumberBENEVISION CENTRAL MONITORING SYSTEM
Catalog Number115-034413-00
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
Manufacturer Address1203 NANHUAN AVENUE GUANGMING GUANGMING DISTRICT SHENZHEN, GUANGDONG 518106 CH 518106


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2018-06-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.