ALGO 2EC NA NO LONGER SELL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2006-08-23 for ALGO 2EC NA NO LONGER SELL manufactured by Natus Medical, Inc..

Event Text Entries

[22076497] The employee was unpackaging the device from the box by holding the foam siding. While removing the device the laptop slid and the employee tried to prevent the laptop from falling to the floor by holding the device by the bottom of the docking station and the employee maneuvered her fingers to better grip the docking station. In this process, her index finger caught between the consol and docking station and her fingernail split down the middle. It appeared that the device was not attached to the docking station. The employee stated that the dual-lock was not fastened correctly to the device and it would have crached to the ground if she had not caught it. The employee stated on 8/17/05 that her finger is "fine" and that she did not seek medical attention.
Patient Sequence No: 1, Text Type: D, B5


[22118700] Natus will implement inspection on devices when packaging to assure that the laptop is securely fastened to the docking station.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2938854-2006-00001
MDR Report Key757767
Report Source05
Date Received2006-08-23
Date of Report2006-08-04
Date of Event2006-07-25
Device Manufacturer Date2000-12-01
Date Added to Maude2006-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLUANNE NG
Manufacturer Street1501 INDUSTRIAL RD
Manufacturer CitySAN CARLOS CA 94070
Manufacturer CountryUS
Manufacturer Postal94070
Manufacturer Phone6508017306
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALGO 2EC
Generic NameHEARING SCREENER
Product CodeEWO
Date Received2006-08-23
Model NumberALGO 2EC
Catalog NumberNA NO LONGER SELL
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key745666
ManufacturerNATUS MEDICAL, INC.
Manufacturer Address1501 INDUSTRIAL RD. SAN CARLOS CA 94070 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-08-23

© 2026 FDA.report