[110362607] Customer service replied with an response, encouraging the customer to make contact via phone so that her issue could appropriately be addressed. In follow up by a complaint handler with the customer on (b)(6) 2018, the customer responded that she started using the gel pads 3 days after delivery for irritated nipples from breastfeeding. She stated that she only wore the gel pads for about 2 hours when she noticed a burning sensation and then saw marks on her breast. She indicated that she stopped using the gel pads, visited her lactation consultant (who consulted her obgyn), and was prescribed a compound steroid cream. The customer indicated that the burns are still noticeable, but are healing. She stated that she has discarded the unused product. Reported issues of rash related to the hydrogel pads were investigated under (b)(4), which determined the probable root causes to be (1) customers with glycerin allergies experiencing aggressive immune reactions due to the fact that glycerin is an integral ingredient in the hydrogel pads and (2) pathogens or allergens that come in contact with exposed hydrogel pads which would subsequently be introduced to the user? S breast.
Patient Sequence No: 1, Text Type: N, H10


[110362608] On (b)(6) 2018, the customer alleged to medela llc via email that the tender care hydrogel pads burned her skin terribly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419937-2018-00141
MDR Report Key7577801
Report SourceCONSUMER
Date Received2018-06-07
Date of Report2018-06-07
Date of Event2018-05-18
Date Mfgr Received2018-05-30
Date Added to Maude2018-06-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAN KLOIBER
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal60050
Manufacturer G1MEDELA LLC
Manufacturer Street1101 CORPORATE DRIVE
Manufacturer CityMCHENRY IL 60050
Manufacturer CountryUS
Manufacturer Postal Code60050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTENDER CARE HYDROGEL PADS
Generic NameNURSING PAD, HYDROGEL
Product CodeNTC
Date Received2018-06-07
Model Number87123
Catalog Number87123
Lot Number051016
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDELA LLC
Manufacturer Address1101 CORPORATE DRIVE MCHENRY IL 60050 US 60050


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-07