STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2018-06-07 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[110376964] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). Investigation results: a stone cone nitinol retrieval coil was received for analysis. A visual analysis of the returned device found the coil was tangled on the end preventing it from closing completely. The coil easily untangled and then opened and closed freely. The blue/green heat shrink is pulled away from the distal stop, the core wire is kinked near the marker bands and the two most distal marker bands show signs of wear. The complaint of the device being unable to close and being tangled, was confirmed due to the condition of the returned device. Since the damage occurred during unpacking, the most probable root cause is? Handling damage". The complaint was caused by handling of the device or portion of the device without direct patient contact either during unpacking, preparation, or shipping; at the end of the procedure; or when packaging for return. A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
Patient Sequence No: 1, Text Type: N, H10


[110376965] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter during a ureteroscopy procedure performed on (b)(6) 2018. According to the complainant, during unpacking, the device was found to be twisted and was unable to close. The patient's condition at the conclusion of the procedure was reported to be? Stable?. This event has been deemed a reportable event based on the investigation results; coil coating peeled/shared/frayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-01784
MDR Report Key7578247
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2018-06-07
Date of Report2018-05-21
Date of Event2018-02-02
Date Mfgr Received2018-05-21
Device Manufacturer Date2017-07-28
Date Added to Maude2018-06-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2018-06-07
Returned To Mfg2018-03-05
Model NumberM0063903200
Catalog Number390-320
Lot Number0003768064
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.