MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-11 for ELECSYS TOTAL PSA IMMUNOASSAY 07027966190 manufactured by Roche Diagnostics.
[110710001]
(b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[110710002]
The customer received questionable elecsys total psa immunoassay and elecsys free psa immunoassay results for one patient sample from cobas e 801 module serial number (b)(4). The medwatchs involved in this event include those with patient identifier = (b)(4). The total psa result was 0. 571 ng/ml and the free psa result was 1. 24 ng/ml. The results were reported outside of the laboratory. There was no allegation of an adverse event. As no sample material was available for further investigation, a specific root cause could not be determined. A general product problem was excluded.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1823260-2018-01827 |
| MDR Report Key | 7586414 |
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
| Date Received | 2018-06-11 |
| Date of Report | 2018-06-11 |
| Date of Event | 2018-04-24 |
| Date Mfgr Received | 2018-05-21 |
| Date Added to Maude | 2018-06-11 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | NA MICHAEL LESLIE |
| Manufacturer Street | 9115 HAGUE ROAD NA |
| Manufacturer City | INDIANAPOLIS IN 46250 |
| Manufacturer Country | US |
| Manufacturer Postal | 46250 |
| Manufacturer Phone | 3175214343 |
| Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
| Manufacturer Street | SANDHOFERSTRASSE 116 NA |
| Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
| Manufacturer Country | GM |
| Manufacturer Postal Code | 68305 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Removal Correction Number | NA |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | ELECSYS TOTAL PSA IMMUNOASSAY |
| Generic Name | TOTAL,PROSTATE SPECIFIC ANTIGEN(NONCOMPLEXED&COMPLEXED) |
| Product Code | MTF |
| Date Received | 2018-06-11 |
| Model Number | NA |
| Catalog Number | 07027966190 |
| Lot Number | 29158500 |
| ID Number | NA |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ROCHE DIAGNOSTICS |
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2018-06-11 |