ELECSYS IGE II IMMUNOASSAY 04827031190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-11 for ELECSYS IGE II IMMUNOASSAY 04827031190 manufactured by Roche Diagnostics.

Event Text Entries

[110708343] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[110708344] The customer received a questionable high elecsys ige ii immunoassay result for one patient sample from modular e module serial number (b)(4). The initial result was 46888 iu/ml (>2500 undiluted) and was reported outside of the laboratory. The results for subsequent electrophoresis testing did not agree with this result as no increased immunoglobulin pattern was visible. Therefore, the sample was sent to another laboratory for confirmation. The result by a siemens method was 1002 iu/ml. The customer repeated an aliquot of the same sample on the modular e module and the results were >2500 undiluted and >50000 with an automatic dilution. There was no allegation of an adverse event. From analysis of the calibration and quality control data, a general reagent issue could be excluded. As no sample material was available, the investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01829
MDR Report Key7586490
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-06-11
Date of Report2018-06-11
Date of Event2018-05-09
Date Mfgr Received2018-05-21
Date Added to Maude2018-06-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS IGE II IMMUNOASSAY
Generic NameRADIOIMMUNOASSAY, IMMUNOGLOBULINS (D,E)
Product CodeJHR
Date Received2018-06-11
Model NumberIGE
Catalog Number04827031190
Lot Number26628902
ID NumberNA
Device Expiration Date2019-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-11

© 2026 FDA.report