STATSPIN SSVT CENTRIFUGE SSVT-1 CENTRIFUGE X00-003917-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-11 for STATSPIN SSVT CENTRIFUGE SSVT-1 CENTRIFUGE X00-003917-001 manufactured by Beckman Coulter.

Event Text Entries

[111164128] Beckman coulter initiated an rma (return material authorization) for replacement of the customer? S centrifuge. The distributor, (b)(4), supplied a new replacement unit to the customer to resolve the issue. Beckman coulter internal identifier is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[111164129] The customer reported the rt12 rotor broke during centrifugation of the statspin ssvt-1 centrifuge. The customer stated that the safety shield was in use at the time of the event. The customer stated that pieces of the rotor had escaped from the ssvt- 1 centrifuge, traveled a few feet and hit a cabinet. There are no reports of harm or injury; thus, no medical attention was received due to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2050012-2018-01001
MDR Report Key7589309
Date Received2018-06-11
Date of Report2018-06-05
Date of Event2018-05-15
Date Mfgr Received2018-05-15
Device Manufacturer Date2011-11-28
Date Added to Maude2018-06-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DAVID DAVIS
Manufacturer Street250 S. KRAEMER BLVD. M/S E1.SE.01
Manufacturer CityBREA CA 928218000
Manufacturer CountryUS
Manufacturer Postal928218000
Manufacturer Phone7149613796
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 928218000
Manufacturer CountryUS
Manufacturer Postal Code928218000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATSPIN SSVT CENTRIFUGE
Generic NameCENTRIFUGES (MICRO, ULTRA, REFRIGERATED) FOR CLINICAL USE
Product CodeJQC
Date Received2018-06-11
Model NumberSSVT-1 CENTRIFUGE
Catalog NumberX00-003917-001
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 928218000 US 928218000


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-11