VITEK? 2 YEAST (YST) IDENTIFICATION (ID) TEST KIT 21343

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-06-12 for VITEK? 2 YEAST (YST) IDENTIFICATION (ID) TEST KIT 21343 manufactured by Biom?rieux, Inc..

Event Text Entries

[111017303] A customer in the united states notified biom? Rieux of a misidentification of cap survey strain candida tropicalis in association with the vitek? 2 yeast (yst) identification (id) test kit. The vitek 2 yst id card identified the strain to candida parapsilosis. Following receipt of the cap answer, the customer retested the original cap strain via yst id card; obtained results of candida lusitaniae and candida haemulonii. Testing of a second specimen from cap provided an organism identification to the expected result of candida tropicalis. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any person's state of health, as there is no patient associated with the cap survey strain, biom? Rieux requested strain submittal from the customer. A biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2018-00222
MDR Report Key7591900
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-06-12
Date of Report2018-07-19
Date Mfgr Received2018-06-26
Device Manufacturer Date2017-08-02
Date Added to Maude2018-06-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JEFF SCANLAN
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318694
Manufacturer G1BIOM
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 YEAST (YST) IDENTIFICATION (ID) TEST KIT
Generic NameVITEK? 2 YST ID CARD
Product CodeJXB
Date Received2018-06-12
Catalog Number21343
Lot Number2430432203
Device Expiration Date2019-02-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX, INC.
Manufacturer Address595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-12