MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-13 for POTASSIUM ELECTRODE COBAS INTEGRA 400/700/800 21029355001 manufactured by Roche Diagnostics.
[111020524]
Unique identifier (udi)#:(b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[111020525]
The customer received questionable high ise indirect gen. 2 potassium results for one patient sample from cobas integra 400 plus serial number (b)(4). The initial results were 6. 37 mmol/l and 6. 54 mmol/l. The result of 6. 37 mmol/l was reported outside of the laboratory. The patient's physician complained about the result and the customer sent the sample to another laboratory that used a cobas 6000 c (501) module. The results were 5. 74 mmol/l and 5. 70 mmol/l. The customer repeated the sample on the cobas integra 400 plus and the results were 5. 64 mmol/l, 7. 29 mmol/l, and 7. 22 mmol/l. There was no allegation of an adverse event. The customer performed precision testing which was acceptable. The customer suspected there was an interference in the sample.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2018-01878 |
MDR Report Key | 7595433 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-06-13 |
Date of Report | 2018-07-19 |
Date of Event | 2018-05-24 |
Date Mfgr Received | 2018-05-28 |
Date Added to Maude | 2018-06-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | AVL SCIENTIFIC CORP. |
Manufacturer Street | 33 MANSELL COURT NA |
Manufacturer City | ROSWELL GA 30076 |
Manufacturer Country | US |
Manufacturer Postal Code | 30076 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | POTASSIUM ELECTRODE COBAS INTEGRA 400/700/800 |
Generic Name | ELECTRODE, ION SPECIFIC, POTASSIUM |
Product Code | CEM |
Date Received | 2018-06-13 |
Model Number | NA |
Catalog Number | 21029355001 |
Lot Number | ASKU |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-06-13 |