POTASSIUM ELECTRODE COBAS INTEGRA 400/700/800 21029355001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-13 for POTASSIUM ELECTRODE COBAS INTEGRA 400/700/800 21029355001 manufactured by Roche Diagnostics.

Event Text Entries

[111020524] Unique identifier (udi)#:(b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[111020525] The customer received questionable high ise indirect gen. 2 potassium results for one patient sample from cobas integra 400 plus serial number (b)(4). The initial results were 6. 37 mmol/l and 6. 54 mmol/l. The result of 6. 37 mmol/l was reported outside of the laboratory. The patient's physician complained about the result and the customer sent the sample to another laboratory that used a cobas 6000 c (501) module. The results were 5. 74 mmol/l and 5. 70 mmol/l. The customer repeated the sample on the cobas integra 400 plus and the results were 5. 64 mmol/l, 7. 29 mmol/l, and 7. 22 mmol/l. There was no allegation of an adverse event. The customer performed precision testing which was acceptable. The customer suspected there was an interference in the sample.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01878
MDR Report Key7595433
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-06-13
Date of Report2018-07-19
Date of Event2018-05-24
Date Mfgr Received2018-05-28
Date Added to Maude2018-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1AVL SCIENTIFIC CORP.
Manufacturer Street33 MANSELL COURT NA
Manufacturer CityROSWELL GA 30076
Manufacturer CountryUS
Manufacturer Postal Code30076
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NamePOTASSIUM ELECTRODE COBAS INTEGRA 400/700/800
Generic NameELECTRODE, ION SPECIFIC, POTASSIUM
Product CodeCEM
Date Received2018-06-13
Model NumberNA
Catalog Number21029355001
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457