NATRELLE INSPIRA X-FULL 800CC SIZER (US) MSZ-X800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-06-13 for NATRELLE INSPIRA X-FULL 800CC SIZER (US) MSZ-X800 manufactured by Allergan (costa Rica).

Event Text Entries

[110954202] Further information from the reporter regarding event, product, or patient details has been requested. No additional information is available at this time. A review of the device history record has been completed. No deviations or non-conformances noted.
Patient Sequence No: 1, Text Type: N, H10


[110954203] Physician reported inspira sizer "ruptured" during surgery. Patient contact with the device occurred. The surgery was successfully completed with a back up device.
Patient Sequence No: 1, Text Type: D, B5


[117903610] Device analysis results: visual analysis of the returned device identified: weight to the spec, crease fold, opening. A microscopic analysis was performed which identify a striated edge opening, consist with use of a surgical tool. Based on the device analysis the final assessment is: a striated edge opening on posterior side due to surgical damage.
Patient Sequence No: 1, Text Type: N, H10


[117903611] Physician reported inspira sizer "ruptured" during surgery. Patient contact with the device occurred. The surgery was successfully completed with a back up device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617229-2018-04474
MDR Report Key7597876
Report SourceHEALTH PROFESSIONAL
Date Received2018-06-13
Date of Report2018-08-15
Date of Event2018-05-16
Date Mfgr Received2018-07-21
Device Manufacturer Date2017-06-01
Date Added to Maude2018-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SUZANNE WOJCIK
Manufacturer Street301 W HOWARD LANE SUITE 100
Manufacturer CityAUSTIN TX 78753
Manufacturer CountryUS
Manufacturer Postal78753
Manufacturer Phone7372473605
Manufacturer G1ALLERGAN (COSTA RICA)
Manufacturer Street900 PARKWAY GLOBAL PARK ZONA FRANCA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNATRELLE INSPIRA X-FULL 800CC SIZER (US)
Generic NameSIZER, MAMMARY, BREAST IMPLANT VOLUME
Product CodeMRD
Date Received2018-06-13
Returned To Mfg2018-07-02
Catalog NumberMSZ-X800
Lot Number3061507
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALLERGAN (COSTA RICA)
Manufacturer Address900 PARKWAY GLOBAL PARK ZONA FRANCA CS


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-06-13