PANORAMIC X-RAY MODEL PC-1000 800724-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-06-13 for PANORAMIC X-RAY MODEL PC-1000 800724-1 manufactured by Young Innovations.

Event Text Entries

[111058098] The dentist emailed and stated that the screw motor assembly failed. The overhead chassis lightly made contact with a girl. The dentist applied antibiotic and a band-aid. The mother took the girl to the physician to make sure she was alright. The physician reapplied antibiotic and a band-aid. The girl was not hospitalized. The dentist purchased a screw motor and had it shipped to his office. He is having his own technician replace the motor since there are no panoramic certified technicians in (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[111058099] Dr. (b)(6) notified us be email on (b)(6) 2018 that three weeks prior (1st week of (b)(6)approximately), a patient was entering the unit and the screw motor assembly failed. The overhead chassis made contact with the patient. On (b)(6) 2018, the doctor gave us the serial number and other required information for this incident so we could investigate and create a case. The screw motor is stuck now and will not go up. The patient was scraped on the cheek and shoulder. The patient was treated with antibiotic and a band-aid on the cheek.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1832462-2018-00003
MDR Report Key7598026
Report SourceHEALTH PROFESSIONAL
Date Received2018-06-13
Date of Report2018-06-13
Date of Event2018-05-02
Date Mfgr Received2018-05-23
Device Manufacturer Date1998-06-23
Date Added to Maude2018-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactADRIENNE DEMLAND
Manufacturer Street4321 GOSHEN RD.
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal46818
Manufacturer Phone8006542027
Manufacturer G1YOUNG INNOVATIONS
Manufacturer Street4321 GOSHEN RD.
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal Code46818
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePANORAMIC X-RAY MODEL PC-1000
Generic NamePC-1000
Product CodeEHD
Date Received2018-06-13
Model Number800724-1
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerYOUNG INNOVATIONS
Manufacturer Address4321 GOSHEN RD. FORT WAYNE IN 46818 US 46818


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-13

© 2024 FDA.report