MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-06-13 for ARCHITECT AFP 03P36-25 manufactured by Abbott Ireland.
[111014770]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. An evaluation is in-process.
Patient Sequence No: 1, Text Type: N, H10
[111014771]
The customer observed a falsely elevated afp result for one patient while using the architect afp assay. The customer provided the following data: on (b)(6) 2018: initial: 10. 62 ng/ml, diluted 1:2: 2. 78 ng/ml, retest of non-diluted specimen: 2. 88 ng/ml there was no adverse impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3008344661-2018-00050 |
MDR Report Key | 7598146 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2018-06-13 |
Date of Report | 2018-07-23 |
Date of Event | 2018-05-21 |
Date Mfgr Received | 2018-07-04 |
Device Manufacturer Date | 2017-11-29 |
Date Added to Maude | 2018-06-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT IRELAND |
Manufacturer Street | DIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK |
Manufacturer City | SLIGO NA |
Manufacturer Country | EI |
Manufacturer Postal Code | NA |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT AFP |
Generic Name | ALPHA-FETOPROTEIN |
Product Code | LOK |
Date Received | 2018-06-13 |
Catalog Number | 03P36-25 |
Lot Number | 81397FN00 |
Device Expiration Date | 2018-10-24 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT IRELAND |
Manufacturer Address | DIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-06-13 |