IPRESARIO SOLO *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-08-25 for IPRESARIO SOLO * manufactured by Del Mar Reynolds Medical, Inc..

Event Text Entries

[503198] The holter analysis system prints strips of holter ecg. Waveforms are displayed correctly; but on the analysis report from which a diagnosis is being made the waveforms are compressed to the point of eliminating the p-wave of the waveform. The p-wave is vital for diagnosing av block. Del mar has attempted to correct the problem stating that the compression of the wave forms was caused by the monitor resolution. A new computer monitor was installed. This did not help the problem. They say they are consulting their engineering section to try to find a software solution. This has been an ongoing issue with their software since its purchase in march of 2006. Follow up reveals: the manufacturer has given the facility a solution deadline of september 31st.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number759886
MDR Report Key759886
Date Received2006-08-25
Date of Report2006-08-25
Date of Event2006-03-23
Report Date2006-08-25
Date Reported to FDA2006-08-25
Date Added to Maude2006-09-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameIPRESARIO SOLO
Generic NameHOLTER SYSTEM, ECG
Product CodeMWJ
Date Received2006-08-25
Model NumberIPRESARIO SOLO
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age11 MO
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key747805
ManufacturerDEL MAR REYNOLDS MEDICAL, INC.
Manufacturer Address13 WHATNEY STREET IRVINE CA 92618 US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-08-25

© 2026 FDA.report