MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-06-14 for BIRD HIGH FLOW BLENDER SERIES 03800A manufactured by Vyaire Medical, Inc.
[111096302]
Vyaire file identification: (b)(4). Any additional information received from the customer will be included in a follow-up report. The suspect device was returned to vyaire's factory service for evaluation. An investigation was performed and was unable to be confirmed as fio2 was in specifications. The o2/air inlet alarm was able to be confirmed as the most likely cause of the issue was due to a faulty alarm cap. The device was repaired and returned to service specifications.
Patient Sequence No: 1, Text Type: N, H10
[111096303]
The customer reported while using the bird high flow blender series; the fraction of inspired oxygen (fio2) is off by twenty percent. The customer reported the o2/air inlet alarms are not working when one of the gases is disconnected. The customer reported there is no patient involvement associated with the event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2021710-2018-07949 |
MDR Report Key | 7601989 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2018-06-14 |
Date of Report | 2018-06-14 |
Date of Event | 2018-05-16 |
Date Mfgr Received | 2018-05-16 |
Device Manufacturer Date | 2013-12-01 |
Date Added to Maude | 2018-06-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | AVERY FOSTER |
Manufacturer Street | 22745 SAVI RANCH PARKWAY |
Manufacturer City | YORBA LINDA CA 92887 |
Manufacturer Country | US |
Manufacturer Postal | 92887 |
Manufacturer G1 | VYAIRE MEDICAL, INC |
Manufacturer Street | 1100 BIRD CENTER DR. |
Manufacturer City | PALM SPRINGS CA 92262 |
Manufacturer Country | US |
Manufacturer Postal Code | 92262 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | BIRD HIGH FLOW BLENDER SERIES |
Generic Name | MIXER, BREATHING GASES, ANESTHESIA INHALATION |
Product Code | BZR |
Date Received | 2018-06-14 |
Returned To Mfg | 2018-06-08 |
Model Number | BLENDER |
Catalog Number | 03800A |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | VYAIRE MEDICAL, INC |
Manufacturer Address | 22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-06-14 |