VISERA ELITE XENON LIGHT SOURCE CLV-S190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-15 for VISERA ELITE XENON LIGHT SOURCE CLV-S190 manufactured by Olympus Medical Systems Corp..

Event Text Entries

[112285652] The subject clv-s190 has not been returned to olympus medical systems corp. (omsc) yet. Omsc will investigate the subject clv-s190 to identify the root cause of this failure phenomenon when omsc receives it. The clv-s190 instruction manual states the corresponding method when there is an abnormality for the device. There were no further details provided. If significant additional information is received, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[112285653] Just before the start of unspecified laparoscopic procedure (before inserting endoscope, after patient anesthesia), the user confirmed that the illumination light was not output from the subject clv-s190 and e101 error (clv temperature error) occurred. Then, the user found that the cooling fan in the subject clv-s190 was not worked. The user replaced the subject clv-s190 with the spare clv-s190, and completed the procedure. There was no report of the patient? S injury regarding this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2018-01162
MDR Report Key7604461
Date Received2018-06-15
Date of Report2018-06-27
Date of Event2018-05-22
Date Mfgr Received2018-06-04
Device Manufacturer Date2011-07-28
Date Added to Maude2018-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KAZUTAKA MATSUMOTO
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone426425177
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVISERA ELITE XENON LIGHT SOURCE
Generic NameXENON LIGHT SOURCE
Product CodeGCT
Date Received2018-06-15
Returned To Mfg2018-05-30
Model NumberCLV-S190
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.