NEUROTHERM? ADAPTOR CABLE NT GENERATOR TO DISPOSABLE DACUK-NT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-06-15 for NEUROTHERM? ADAPTOR CABLE NT GENERATOR TO DISPOSABLE DACUK-NT manufactured by Neurotherm, Inc.

Event Text Entries

[111151506] The results of the investigation are inconclusive since the device was not returned for analysis. Review of the device history record was not possible as the lot number is unknown. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[111151507] Related manufacturer reference number: (b)(4). During a procedure port 1 on the generator displayed an unsafe probe error message when going into lesion mode. The procedure was cancelled and there were no adverse consequences to the patient. Further troubleshooting determined the issue was due to two generator adapter cables.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002953813-2018-00036
MDR Report Key7604827
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-06-15
Date of Report2018-06-15
Date of Event2018-05-21
Date Mfgr Received2018-05-22
Date Added to Maude2018-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDENISE JOHNSON
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517564470
Manufacturer G1NEUROTHERM, INC
Manufacturer Street600 RESEARCH DRIVE STE 1
Manufacturer CityWILMINGTON MA 01887
Manufacturer CountryUS
Manufacturer Postal Code01887
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEUROTHERM? ADAPTOR CABLE NT GENERATOR TO DISPOSABLE
Generic NameMEDICAL DEVICE ELECTRICAL CABLE, REUSABLE
Product CodeIKD
Date Received2018-06-15
Model NumberDACUK-NT
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNEUROTHERM, INC
Manufacturer Address600 RESEARCH DRIVE STE 1 WILMINGTON MA 01887 US 01887


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-15