SCULPSURE 100-7026-010 105-7026-000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2018-06-15 for SCULPSURE 100-7026-010 105-7026-000 manufactured by Cynosure Inc..

Event Text Entries

[111165049] There was no medication or intervention associated with this event. User error was involved since the operator performed the sculpsure laser treatment over an existing tattoo near the hip bone, which is cautioned against in the clinical reference guide. This resulted in the tattoo area receiving a burn that may develop into a potential scar. Per the sculpsure clinical reference guide: "when treating over a tattoo, there is an increased risk of adverse skin effects and discomfort if the applicator(s) is not in full contact with the skin" and "do not treat over scars or bony area such as hips or ribs. Discomfort or adverse skin effects may occur". As a result, there was a loss of proper contact with the skin during the treatment and prevented adequate cooling from being delivered to the treated areas. The device was not evaluated due to the user error in this incident. Burns are expected side effects from laser treatments, however this is reportable because of the potential for scarring.
Patient Sequence No: 1, Text Type: N, H10


[111165050] Patient experienced a burn on the flanks area following a laser procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222993-2018-00010
MDR Report Key7605085
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2018-06-15
Date of Report2018-06-15
Date of Event2018-05-10
Date Mfgr Received2018-05-18
Device Manufacturer Date2016-12-29
Date Added to Maude2018-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street5 CARLISLE ROAD
Manufacturer CityWESTFORD MA 01886
Manufacturer CountryUS
Manufacturer Postal01886
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSCULPSURE
Generic NameSCULPSURE
Product CodePKT
Date Received2018-06-15
Model Number100-7026-010
Catalog Number105-7026-000
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCYNOSURE INC.
Manufacturer Address5 CARLISLE ROAD WESTFORD MA 01886 US 01886


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.