BCS XP SYSTEM 10461894

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-15 for BCS XP SYSTEM 10461894 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[111489988] The customer contacted siemens customer care center (ccc) and reported that discordant "no clot" results were obtained on multiple patient samples, without further flags. A siemens field service engineer (fse) was dispatched to the customer's site. Siemens is investigating the issue.
Patient Sequence No: 1, Text Type: N, H10


[111489989] The customer reported that discordant activated partial thromboplastin time (aptt) results of "no clot" were obtained on multiple patient samples on the bcs xp system when the pathromtin sl reagent vial was empty. The discordant results were not reported to the physician(s). The samples were rerun on an alternate bcs xp system or on the same bcs xp system after replacing the reagent vial, resulting in correct aptt results. The correct aptt results were reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant "no clot" aptt results obtained on patient samples.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2018-00052
MDR Report Key7607975
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-06-15
Date of Report2018-06-22
Date of Event2018-05-22
Date Mfgr Received2018-05-28
Date Added to Maude2018-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINA LAM
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243504
Manufacturer G1BIT ANALYTICAL INSTRUMENTS GMBH
Manufacturer StreetAM KRONBERGER HANG 3 REGISTRATION NUMBER:3003601075
Manufacturer CitySCHWALBACH, 65824
Manufacturer CountryGM
Manufacturer Postal Code65824
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBCS XP SYSTEM
Generic NameBCS XP SYSTEM
Product CodeGKP
Date Received2018-06-15
Model NumberBCS XP SYSTEM
Catalog Number10461894
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL-VON-BEHRING-STR. 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.