ETEST? ITRACONAZOLE 525858

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-06-20 for ETEST? ITRACONAZOLE 525858 manufactured by Biomerieux Sa.

Event Text Entries

[112571794] A customer in the united states notified biom? Rieux of out of range low results for internal quality controls (qc) in association with etest? Itraconazole (reference 525858). The customer reported that at 24 hours they had a mic value of 0. 016 on multiple test strips. At 48 hours the customer reported test results of 0. 032. The customer has tested the same lot of strips with multiple lots of organisms and media. The customer ordered another order of strips, and received the same lot. They tested the new kit of etest? Strips and received similar results. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00086
MDR Report Key7620578
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-06-20
Date of Report2018-08-20
Date Mfgr Received2018-07-24
Device Manufacturer Date2017-02-16
Date Added to Maude2018-06-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer Street3 ROUTE DE PORT MICHAUD
Manufacturer CityLA BALME LES GROTTES ISERE 383,
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? ITRACONAZOLE
Generic NameETEST? ITRACONAZOLE
Product CodeNGZ
Date Received2018-06-20
Catalog Number525858
Lot Number1005507630
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383, FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-20

© 2026 FDA.report