CAVIWAVE ULTRA SONIC CLEANER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-20 for CAVIWAVE ULTRA SONIC CLEANER manufactured by Steris Corporation - Distribution Center.

Event Text Entries

[112090967] A steris service technician arrived onsite following the reported event and spoke with user facility personnel. User facility personnel stated that they observed a small "flame" and smoke emitting from the unit's drain piping. Immediately prior to the event, the unit successfully completed a cycle and was not in use during the time of the reported event. The steris technician inspected the unit and found that the transducer cable was damaged. The transducer cable provides electrical power from the ultrasonic generator to the transducers which are bonded to the underside of the caviwave cleaner's wash tank. During the last routine preventive maintenance in march 2018, no issues were noted with the transducer cable; the cleaner was found to be operating according to specification. A 3-year review of complaint records indicates this to be an isolated event. Due to the damage sustained when extinguishing the flame, steris is working with the user facility to provide a replacement unit.
Patient Sequence No: 1, Text Type: N, H10


[112090968] The user facility reported that a small flame was observed from the caviwave ultra sonic cleaner. User facility personnel immediately called the on-duty fire fighter. The on-duty fire fighter quickly extinguished the flame. No injuries were associated with the reported event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003950207-2018-00004
MDR Report Key7622742
Date Received2018-06-20
Date of Report2018-06-20
Date of Event2018-05-19
Date Mfgr Received2018-05-22
Date Added to Maude2018-06-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS CORPORATION - DISTRIBUTION CENTER
Manufacturer Street6100 HEISLEY RD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal Code44060
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCAVIWAVE ULTRA SONIC CLEANER
Generic NameCAVIWAVE CLEANER
Product CodeFLG
Date Received2018-06-20
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - DISTRIBUTION CENTER
Manufacturer Address6100 HEISLEY RD MENTOR OH 44060 US 44060


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-20