MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-06-21 for VITEK? 2 GN TEST KIT 21341 manufactured by Biomerieux, Inc..
[112566087]
A customer in canada reported a misidentification of salmonella enterica as shigella species in association with the vitek 2 gn test kit (lot 2410196103). The gn card identified the strain as shigella group (88%), but it was also identified as salmonella enterica by pcr and a reference laboratory. The customer did not provide patient information. There is no indication or report to biom? Rieux that the discrepant result led to any adverse event related to a patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1950204-2018-00232 |
| MDR Report Key | 7625612 |
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
| Date Received | 2018-06-21 |
| Date of Report | 2018-09-18 |
| Date Mfgr Received | 2018-08-21 |
| Device Manufacturer Date | 2017-06-10 |
| Date Added to Maude | 2018-06-21 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | MS. DEBRA BROYLES |
| Manufacturer Street | 595 ANGLUM ROAD |
| Manufacturer City | ST. LOUIS MO 63042 |
| Manufacturer Country | US |
| Manufacturer Postal | 63042 |
| Manufacturer Phone | 3147318805 |
| Manufacturer G1 | BIOMERIEUX, INC. |
| Manufacturer Street | 595 ANGLUM ROAD |
| Manufacturer City | HAZELWOOD MO 63042 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 63042 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | VITEK? 2 GN TEST KIT |
| Generic Name | VITEK? 2 GN TEST KIT |
| Product Code | LQM |
| Date Received | 2018-06-21 |
| Catalog Number | 21341 |
| Lot Number | 2410196103 |
| Device Expiration Date | 2018-06-10 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | N |
| Device Age | DA |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | BIOMERIEUX, INC. |
| Manufacturer Address | 595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2018-06-21 |