SADDLELOOP 18G X 10CM WITH BLUNT NEEDLE 1541

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-06-22 for SADDLELOOP 18G X 10CM WITH BLUNT NEEDLE 1541 manufactured by Quest Medical, Inc.

Event Text Entries

[112573630] The alleged issue occurred outside of the us. A similar device is distributed by quest medical in the us. The device is expected to be returned and an investigation will be completed when the device is received. A follow up medwatch will be submitted if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


[112573631] A report was received regarding an alleged issue with the device. The report states that when the hospital staff opened the device package, they observed that the tensioner on the saddleloop was almost detached from the saddleloop. There were no patient complications resulting from the alleged issue.
Patient Sequence No: 1, Text Type: D, B5


[132403611] Investigation conducted showed inconsistencies in dispensing the adhesive which holds the components of the device together. The dhr review showed the device was however within specification having been subjected to a maximum tensioner break force of >1. 0 ibs without failing. Quest medical has implemented a process change as a proactive measure to prevent this issue. The process improvement involves a change in the needle size of the adhesive dispensing gauge for consistent adhesive dispensing.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1649914-2018-00062
MDR Report Key7629642
Date Received2018-06-22
Date of Report2018-01-17
Date Mfgr Received2018-06-15
Date Added to Maude2018-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTOSAN ONOSODE
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal75002
Manufacturer Phone9723326338
Manufacturer G1QUEST MEDICAL, INC
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal Code75002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSADDLELOOP 18G X 10CM WITH BLUNT NEEDLE
Generic NameVASCULAR LOOP
Product CodeGAE
Date Received2018-06-22
Model Number1541
Lot Number049402
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerQUEST MEDICAL, INC
Manufacturer AddressONE ALLENTOWN PARKWAY ALLEN TX 75002 US 75002


Patients

Patient NumberTreatmentOutcomeDate
10 2018-06-22

© 2026 FDA.report