ROHO? UNIVERSAL CUSHION COVER UCFR79

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2018-06-22 for ROHO? UNIVERSAL CUSHION COVER UCFR79 manufactured by Roho, Inc..

Event Text Entries

[112066096] The customer did not state that the cushion cover malfunctioned. The customer also stated that 23 hours of continual use was not normal use. While the customer reported a serious injury, roho has not seen medical records to confirm the injury. Roho has sent the customer a new cover and included a return label for the cover in use at the time of the injury, so we can evaluate the product. If the cover is returned, roho will submit a follow-up report after it has been evaluated. This report was initially submitted on 11/3/2017. At that time, acknowledgements were received stating that the submission was received. However, after some months now, it has become evident that the submission failed due to a typo in the manufacturer's report number. Instead of "1415709", it should have been "1419507". [(b)(4)].
Patient Sequence No: 1, Text Type: N, H10


[112066097] The customer stated he had a pressure ulcer for the first time in 29 years. He stated he received a new cushion cover, dated 2015, and that he noticed the top material of the cover was different from his previous covers when he dropped a cigarette on it and the material did not burn. After roho contacted the customer for additional information, he stated that he had been in the hospital and/or on bed rest for 3 months. He also stated his injury was on his left ischial and had required surgery. He told us he noticed the pressure injury after he sat on the cushion for 23 hours straight, which was not typical use for him. He noticed his bottom and his blue jeans were sweatier than usual after this prolonged, continual use of the cushion and cover and this may have contributed to the injury. He did not think that cushion or cover malfunctioned, but thought that the different cover material did not have the same airflow as the previous material.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419507-2017-00006
MDR Report Key7630381
Report SourceCONSUMER
Date Received2018-06-22
Date of Report2017-11-03
Date of Event2017-08-01
Date Mfgr Received2017-10-05
Device Manufacturer Date2015-01-01
Date Added to Maude2018-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ROBIN GERGEN
Manufacturer Street100 N. FLORIDA AVE.
Manufacturer CityBELLEVILLE IL 62221
Manufacturer CountryUS
Manufacturer Postal62221
Manufacturer Phone8008513449
Manufacturer G1ROHO, INC.
Manufacturer Street1501 S. 74TH STREET
Manufacturer CityBELLEVILLE IL 62223
Manufacturer CountryUS
Manufacturer Postal Code62223
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROHO? UNIVERSAL CUSHION COVER
Generic NameWHEELCHAIR CUSHION COVER
Product CodeKNO
Date Received2018-06-22
Model NumberUCFR79
Lot Number2015
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROHO, INC.
Manufacturer AddressROHO, INC. 1501 S. 74TH STREET BELLEVILLE IL 62223 US 62223


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2018-06-22

© 2024 FDA.report